Clarify. TPS Reset Failure. 2249118 SRI 2 Hardware Configuration Conflict

Symptom

Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis

Error message

2224839 SRI Software Error. A TPS Reset may be necessary. The SRI Extended System could not be started due to an error in the control block.

2249110 SRI Communication Error. SRI failed to respond to a packet from the SCP. First packet opcode waiting for a response: 0x47

2249118 SRI 2 Hardware Configuration Conflict. The SRI 2 System Configuration Dip Switch is set to 0x1, but should be set to 3 (0011 binary) for a Cylindrical System

2223817 TPS is Not Responding. Please Reset TPS Wed Apr 28 16:58:01 2004 Error: 2223817 Host: GEMS Process: NSP File: TardisReset.cc Line: 398 TPS is not responding. Please reset TPS.

2249113 SRI Configuration Failed. SRI Link Errors Caused the Configuration to Fail. See Long Description Wed Apr 28 16:58:01 2004 Error: 2249113 Host: GEMS Process: NSP File: SCP:SriAutoConfig Line: 384 SRI Configuration Failed. SRI link errors caused the configuration to fail. Please fix and perform a TPS Reset.

2224829 SRI-SCP Communication Error. Wed Apr 28 16:58:01 2004 Error: 2224829 Host: GEMS Process: NSP File: SCP:SRI Line: 200 SRI-SCP Communication Error. A TPS Reset may be necessary. The SRI received a packet from the SCP containing an unknown opcode.

Applies to

Models
Excite
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Area
MR

Full cause & step-by-step resolution