Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT182407

Table failed to move up or down with error 182407

Symptom

Table will not move up or down during a elevation table reference. Table will not move up or down during a elevation table reference

Error message

Error: 182407 Velocity error detected when trying to find the Index Pulse Sun Aug 29 13:25:59 2004 Error: 182407 Host: ETC Process: Elevation File: CmdMotion.c 1.35 12/23/93 Line: 174 Function: PATIENT ANATOMY POSITIONING : Elevation - Tilt Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 13:25:59:690 Log Series: 34 HW/SW error: Velocity error detected when trying to find the Index Pulse. Hardware Inhibit Switches: 0xc0d8 Elev Direction: 0 (0 = Up 1 = Down). INFO: Probable Problem could be 180V Power is LOW.

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 2.X (Plus, QX/i) - GRE always an SDAS 4 slice H2_P_S4 (SDAS4 and H2 gantry) Lightspeed 2.X - GRE Method of visual identification Small rounded gantry similar in appearance to LS 3.X and 4.X. Always has an SDAS 4 slice Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) -h or –help = Outputs complete details about all the available switches associated with swhwinfo swhwinfo –c –m –o -da List of Supported Console Types GOC3 Will have a DARC, SDDA and IG GOC4 Will have a DARC 2 and IG List of softwares the product can run 07MW11.10 SP3 06MW29.7 - obsolete 06MW03.5 - obsolete 05MW14.5 - obsolete 04MW44.9 - obsolete 04MW13.7 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-288170G1 ELEV/TILT AMP See LD for more 46-288170G1 ELEV/TILT AMP Part #: 46-288170G1 Part Type: GP CONSUMABLES Part Weight: 1.36 LBR Repairable: No Returnable: No Hazardous Material: N Description: ELEV/TILT AMP FRU: Yes Status: Orderable Comments: FLS - B: THE PROBLEM IS USUALLY WITHTHE 24 V RELAY. PLEASE HAVE THE FE TRY PART 2120791-9. MSA6521 5/29/01 DELISTED T3202CF 12/00 MS5277 FLS - PLEASE CONVERT SPH ORDER TO OVERNIGHT SHIPMENT IF PART AVAILABLE IN MEMPHIS AND IF FEDEX SERVICE (INCLUDING 8:00 AM SERVICE) MEETS FE NEED TIME CRITERIA FE ASSIST: IN ORDER TO INCREASE REVENUE GLOBAL PARTS HAS CREATED A PROGRAM TO HARVEST PARTS FROM RENEWABLE RESOURCES. THESE PARTS OFFER DIRECT SAVINGS TO THE BUSINESSAND THE FIELD
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues