GE Healthcare · CTLightSpeed18328950001
LightSpeed 3.X (Ultra, Plus, QX/i) -Octane
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 183289 Unexpected Loss Of High Voltage Wed Aug 4 20:21:15 2004 Error: 183289 Host: OBC Process: Generator File: gen xray.c 1.31 Line: 528 Function: XRAY GENERATION : Rotating HVON Control Scan: 50001/1/1 Type: Axial Exception Level: Pri/Most Severe Time: 20:21:16:325 Log Series: 59 Unexpected Loss Of High Voltage. .
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-296221P15 80A 600V Fuse See Long Description for details Part #: 46-296221P15 Part Type: GP CONSUMABLES Part Weight: 0.13 LBR Repairable: No Returnable: No Hazardous Material: N Description: BUSSMANN KBC80 FRU: Yes Status: Orderable Comments: FLS - CLASS 1 FRU. PDV BUSSMANN KBC80 80A, 600V THERE ARE 3 OF THESE FUSES IN FUSE KIT 2218570 - MS4751 8/01 PRICE CHANGE PER BRIAN NAGLE 4/98 FLS - FE ASSIST: IN ORDER TO INCREAS INCREASE REVENUE GLOBAL PARTS HAS CREATED A PROGRAM TO HARVEST PARTS FROM RENEWABLE RESOURCRESOURCES. THESE PARTS OFFER DIRECT SAVINGS TO THE BUSINESSAND THE FIELD. ORDER THE U PART WHEN AVAILABLE. IF NO U USED PARTS ARE AVAILABLE CARES WILL CHANGE THE ORDER TOA NEW PART.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required