GE Healthcare · MRExcite2252239
2252239 Startup Timeout was detected on the Universal Power Monitor due to UPM Processor Board.
Symptom
Fails TPS Reset when booting up to Applications or when Resetting TPS See Long Description When the MR system is booting up to applications it will fail TPS reset but Applications will successfully boot. With Applications up and "Reset TPS" has been selected, TPS fails to download successfully.
Error message
2252239 Startup Timeout was detected on the Universal Power Monitor 1. Fri Jul 15 12:26:57 2005 2252239 2 lx-mr NSP SCP:CanAutoConfig 2576 Startup Timeout was detected on the Universal Power Monitor 1.-> 5600 The ASC NB and/or ASC BB IF Boards were detected but the UPM1 and/or UPM2 Processor Boards were not 5600 Process: msgView Host: MR30F037 Process: NSP File: SCP:CanAutoConfig Line: 2983 Exception Class: Unknown Severity: Unknown Function: Software/Firmware/Config: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 The ASC NB and/or ASC BB IF Boards were detected but the UPM1 and/or UPM2 Processor Boards were not. Verify the RF Amp/UPM configuration and perform a TPS Reset 5600 Process: msgView Host: MR30F037 Process: NSP File: SCP:CanAutoConfig Line: 2983 Exception Class: Unknown Severity: Unknown Function: Software/Firmware/Config: N/A PSD: Unknown Coil: Unknown Scan: -1/-1 The ASC NB and/or ASC BB IF Boards were detected but the UPM1 and/or UPM2 Processor Boards were not. Verify the RF Amp/UPM configuration and perform a TPS Reset.
Applies to
- Models
- Excite
- Product
- Excite HDx Includes "Excite HDi" system. 14.0 Software only.
- Software
- 14.0 For both Excite HDx and 3T Excite HDx systems.
- Component
- 2359518-2 UPM PROCESSOR BOARD.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.