Unable to start Xray Scan or QC, getting message failed to start sequence
Symptom
Unable to start Xray Scan or QC, getting message failed to start sequence Unable to start Xray Scan or QC, getting message failed to start sequence.
Error message
ERROR_SEQUENCE_CONNECT (ERROR 11404) ERROR_SEQUENCE_START:.:Failed to start sequence.:.:1. Close session 2.Try again 3. Partialy shutdown the application :.: Failed to start sequence Static error in operation: connect.:.:Trace:Caused by:Failed to start sequence. (11411 ERROR) Failed to connect to the scanner (11404 ERROR) The Application has problems to communicate with one of the servers. (9004 ERROR) Failed to perform a subsystem service. SubSystem=feServer; Service=connectServer; Param=null (9501 ERROR) Failed to send a command to one of the scanner servers (10003 ERROR) Empty input buffer; could not read buffer length. ERROR=Error in GantryServer d:\tacq\su\src\cpp\gantry\server\stretcheraxis\wrapdriver.cpp(134):: Error number => 65 = Disabled by SW. Motion could not begin because a switch programmed to abort motion was active when MO=1 was tried.
Applies to
- Models
- Infinia
- Product
- Infinia HE4 Infinia II Hawkeye 4 & Infinia II GP (Infinia running Windows XP software) Infinia II Infinia II Acquisition System Infinia I Infinia I Acquisition System VGH VG and VG Hawkey and Varicam
- Component
- *X-Ray (Generator, DAS, Controller) *X-Ray (Generator, DAS, Controller)
- Area
- Nuclear Medicine
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare Nuclear Medicine issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare Nuclear Medicine issues
- Service Note 43783_1 Feb/08 Infinia/Ventri/Xeleris 1.1 & 2
- After software upgrade FMI40786, the Infinia application does not automatically start on boot up
- I need the product data sheet on catalog E4500AK x-ray in use warning light interface control panel for Infinia
- Infinia Does not recognize collimator is mounted to detector, not reading collimator ID correctly.
- Where do I obtain the Non-Linear Correction tables for a VG/Infinia Hawkeye system if can't find the CD that came with the DAS assembly?
- Hawkeye CT scan Ring (Circular) artifacts in X-ray images appearing in the same radius in all of the images.