Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT182829

LightSpeed 4.X - Octane

Symptom

Customer was not able to remote tilt gantry. In many cases , gantry will tilt from the Gantry Controls.

Error message

Error: 182829 Remote Tilt Input to the ETC I/F Board was in an INVALID State , Remote Tilt Functionality is Disabled until ETC I/F is Reset See LD for the full error Mon Aug 8 10:35:43 2005 Error: 182829 Host: ETC Process: Gantry Display File: gd_tnc_erlog.c Line: 651 Function: No System Function Reported Scan: 1111/2/0 Type: None/Unknown Exception Level: Pri/Soft Time: 10:35:44:730 Log Series: 179 Remote Tilt Input to the ETC I/F Board was in an INVALID State. Remote Tilt Functionality is Disabled until ETC I/F is Reset. Check Remote Tilt Cabling.

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues