GE Healthcare · CTLightSpeed2300090671098923000653025088000000800000
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane
Symptom
Image reconstruction failures. Image reconstruction failures
Error message
Error: 230009067 Retry queue entry See Long Description for error example. Tue Oct 19 09:24:04 2004 Error: 230009067 Host: dgct Process: queuemgr File: qmgr_list_mgmt.c Line: 2217 Function: Image Generation : SBC Processing Job Type: Prospective Helical Scan: 10989/2/1 Image: 10989/2/1 Retry queue entry Error: 230006530 RECON_JOB_FAILED_RETRY See Long Description for complete error Tue Feb 1 06:33:34 2005 Error: 230006530 Host: ice Process: reconmgr File: rm_recon_mgr.c Line: 1114 Function: Image Generation : SBC Processing Job Type: Prospective Axial Scan: 25088/2/9 Image: 25088/3/25 Recon Seq Id: 957 RECON_JOB_FAILED_RETRY : number_detector_macro_rows 4 collimation_mode 0 num _rows_used_per_image 1 vw_mode -1 zsmooth_factor 1.000000 ibo 1 tilt 0.000000 gantry_period_sec_rev 0.800000 Error: 230005024 image recon seq id: 18997Rebin failed in row delay, status=2 See Long Description for complete error Mon Feb 6 12:12:50 2006 Error: 230005024 Host: ice-post Process: sdcctrl File: sdcctrl_events.cxx Line: 447 image recon seq id: 18997Rebin failed in row delay, status=2 Error: 230006529 Recon Mgr is sending queue ack to queuemgr with a RECON_JOB_FAILED_RETRY status See Long Descript
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2225004-9 73GB scan disk (replaces 2225004-8 and 18 GB p/n 2225004-4 and 36 GB scan disk 2225004-5) See Long Description for Details. Part #: 2225004-9 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: 73GB 15000 RPM Scan Data Drive FRU: Yes Status: Orderable Comments: Per ECR 2046173, 212008898, 8/8/07. ** FE ASSIST ** Refer to Service Note SN07110612a For Jumper information (12/3/07 212001261) Part #: 2225004-8 Part Type: GP CONSUMABLES Part Weight: 1.61 LBR Repairable: No Returnable: No Hazardous Material: N Description: HARD DRIVE 73GB 15KRPM SCAN DATA FRU: Yes Status: Orderable Comments: Added per ECR 2031409, 7/14/06, ms11892. ** FE ASSIST ** If no stock of the 2225004-8 order 2225004-9 as the replacement. Refer to Service Note SN07110612a (212001261 12/3/07)
- Area
- CT
Full cause & step-by-step resolution