Pular para o conteúdo principal
GE Healthcare · MRSigna2213312

Unable to scan, Erbtec errors.

Symptom

Unable to scan with Erbtec errors

Error message

2213312 ERBTEC 20KW RF Amplifier fault 85 RF Power Monitor Fault ERROR INFO: 14:15:47.....23-Sep-97.........................Error Code: 2213312 STAT 00 in [ ST MONITOR] Abcode: 0 /srv ERBTEC 20KW RF Amplifier fault 85. RF power monitor fault. The 12v externally controlled HV safety relay on the Processor Board was disabled by the GE RF power monitor. Possible defective GE power monitor, cables, or Processor Board.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues