GE Healthcare · MRExcite2254204
Will Not Scan With RF Errors.
Symptom
Will Not Scan With RF Errors Will not scan with RF errors
Error message
2254204 SRFD2 RF Amplifier Power Supply Fault. RF Amplifier indicates the High Voltage Power Supply is out of Specification. Fri Jun 25 13:04:45 2004 Error: 2254204 Host: lx-mr Process: NSP File: HPC_PORT_ID:spi_rf_mon Line: 1717 SRFD2 RF Amplifier Power Supply Fault. RF Amplifier indicates the high voltage power supply is out of specificati on.sible causes: Supply feedback loop exceeds 2.5v, PS heat sink is overheated,supply FETS are shorted. Extended Error: Check fans, may be blocked or defective, Cabinet Filters need to be changed, or RF Amplifier may be operating out of specification.
Applies to
- Models
- Excite
- Product
- Excite HDx Includes "Excite HDi" system. 14.0 Software only.
- Software
- 14.0 For both Excite HDx and 3T Excite HDx systems.
- Area
- MR
Full cause & step-by-step resolution