GE Healthcare · CTLightSpeedVCT18242615403
Lightspeed Ultra 3.x
Symptom
Customer was not able to remote tilt gantry. In many cases , gantry will tilt from the Gantry Controls.
Error message
Error: 182426: Gantry Touch Sensor FAULT. Console Pushbutton Remote Tilt is DISABLED See Long Description for Full Error Tue Jan 10 17:42:00 2006 Error: 182426 Host: ETC Process: Elevation File: emcTouchSensor.c Line: 97 Function: PATIENT ANATOMY POSITIONING : Elevation - Tilt Scan: 7855/2/0 Type: None/Unknown Exception Level: Pri/Soft Time: 17:41:59:940 Log Series: 197 The Gantry Touch Sensor FAULT. Console Pushbutton Remote Tilt is DISABLED. (Gantry Sensors - FORWARD Touch Sensor Error(s)). Gantry Controls are running in FAULT MODE. Current Position 15403 (.01 mm). Elev Dir: 1 (0 = Up 1 = Down). (Elev Cmd: 1 TouchSensor: 0x5000).
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2277405 Front / Small Gantry Tilt Forward Sensor / Interference Switch used on H2 gantry See Long Description for Details 2277405 Front / Small Gantry Tilt FORWARD Sensor / Interference Switch used on H2 Part #: 2277405 Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: PATIENT SENSOR, SMALL FRU: Yes Status: Orderable Comments : FLS - SAFETY TILT INTERFERENCE SWITCH FRONT USED ON H2 COMMON TO MFG. MS26561 DSM 1/7/01 COLLECTION OF DEFECTIVES START7/23/01 MS5277 FLS - PLEASE CONVERT SPH ORDER TO OVERNIGHT SHIPMENT IF PART AVAILABLE IN MEMPHIS AND IF FEDEX SERVICE (INCLUDING 8:00 AM SERVICE) MEETS FE NEED TIME CRITERIA
- Area
- CT
Full cause & step-by-step resolution