GE Healthcare · MRSigna2225518
The center conductor of the Quad Head Coil cable 46-282441P2 (white ends) was found shorted to shield.
Symptom
Cannot Scan with Head T/R Driver Error. The system failed to scan either head or body scans. The SSM light indicated T/R Fault.
Error message
2225518 Dynamic disable and T/R driver fault. Possibly a shorted diode in the head T/R switch network, or a short circuit load. See long description Thu Feb 6 04:02:38 1997 2225518 1 lx-bay14b NSP HPC_PORT_ID:spi_tr_mon 6060 Dynamic disable and T/R driver fault. Possibly a shorted diode in the head T/R switch network, or a short circuit load.
Applies to
- Models
- Signa
- Product
- Signa LX Signa LX Product can be used with 8.x and 9.x systems.
- Software
- 9.1 For Signa LX product only
- Component
- 46-282441P2 Quad Head Coil cable white ends 46-282441P1 BLUE ENDS CABLE 46–282441P1 BLUE ENDS CABLE 46–282441P2 WHITE ENDS CABLE 46-282441P2 Quad Head Coil cable white ends
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.