GE Healthcare · MRSigna2225518

The center conductor of the Quad Head Coil cable 46-282441P2 (white ends) was found shorted to shield.

Symptom

Cannot Scan with Head T/R Driver Error. The system failed to scan either head or body scans. The SSM light indicated T/R Fault.

Error message

2225518 Dynamic disable and T/R driver fault. Possibly a shorted diode in the head T/R switch network, or a short circuit load. See long description Thu Feb 6 04:02:38 1997 2225518 1 lx-bay14b NSP HPC_PORT_ID:spi_tr_mon 6060 Dynamic disable and T/R driver fault. Possibly a shorted diode in the head T/R switch network, or a short circuit load.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
46-282441P2 Quad Head Coil cable white ends 46-282441P1 BLUE ENDS CABLE 46–282441P1 BLUE ENDS CABLE 46–282441P2 WHITE ENDS CABLE 46-282441P2 Quad Head Coil cable white ends
Area
MR

Full cause & step-by-step resolution