CWA ID83324. Mouse and Keyboard Quit Working.

Symptom

Mouse and Keyboard Quit Working See Long Description

Error message

2225382 - SPI-Workspace Interface Module Communication Error Error: 2225382 Host: SMRITR1 Process: NSP File: HPC_PORT_ID:spi_cpb_wdog_fail Line: 3 SPI-Workspace Interface Module Communication Error. The SPI link to the Workspace Interface Module failed - failure detected via timeout. DUART status: 0x0C Please Reset TPS. If error is persistent, disconnect and reconnect the power cable to the WIM. If error still exists, replace the WIM.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR