Pular para o conteúdo principal
GE Healthcare · MRSigna

Octane fan failed causing overheating.

Symptom

Octane Computer Panic Kernel Error Fault The SGI Octane computer won't boot to the login screen and shows panic kernel bus faults and codes on the host monitor. Light bar may or may not go from white to red and back to white again. Computer crashes, performs a core dump, and displays panic kernel error.

Error message

PANIC Kernel Bus fault

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
2138601-8 Fan Module Assembly 2138601-8 Fan Module Assembly 2138601-8 Fan Module Assembly