Pular para o conteúdo principal
GE Healthcare · MRExcite2248650

Will Not Scan With Patient Comfort Sensor Trip.

Symptom

Will Not Scan With Patient Comfort Sensor Trip. See Long Description Patient comfort sensor trip; cannot scan due to bore sensor indicating temperature is too hot for patient safety. A message "Patient Comfort Sensor Trip" appears in the message box on the left side of the scan page at the start of autoprescan.

Error message

2248650 The patient comfort level is warmer than normal. The patient may be uncomfortable. Further increases in the temperature will inhibit scanning. 2248650 b4a NSP SCP:BoreTempSriHWO 987 The patient comfort level is warmer than normal. The patient may be uncomfortable. Further increases in the temperature will inhibit scanning. Verify: 1. Patient fan is ON. 2. Room temperature is normal. 3. Air flow through the bore is unobstructed.

Applies to

Models
Excite
Product
Excite HDx Includes "Excite HDi" system. 14.0 Software only.
Software
14.0 For both Excite HDx and 3T Excite HDx systems.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues