GE Healthcare · MRExciteOvation2248356
CWA ID82468. Manual Pre Scan MPS runs for 5 to 10 seconds and then stops with error SSI Watchdog Timeout.
Symptom
Manual Pre Scan MPS runs for 5 to 10 seconds and then stops with error: SSI Watchdog Timeout. New Excite II system install running 11.0 M4 software. The FE is trying to do the RF cals. Manual Prescan runs for 5 to 10 seconds and then stops with error: "SSI Watchdog Timeout. An SSI was not detected in the timeout period. The PSD or AGP might not be running.
Error message
2248356 SSI Watchdog Timeout. An SSI Was Not Detected in the Timeout Period. See Long Description Wed Jul 7 09:33:17 2004 Error: 2248356 Host: GEMRI_1Process: NSP File: SCP:sct_wdog_exp Line: 852 SSI Watchdog Timeout. An SSI was not detected in the timeout period. The PSD or AGP might not be running. It may be necessary to perform a TPS reset. If TPS Reset doesn't help then AC 60Hz clock may be missing. Extended Error: Check the 60Hz clock signal from PDM to STIF Board in CAM chassis. Run CAM Data Acquisition SRF Diagnostics.
Applies to
- Models
- Excite, Ovation
- Product
- Excite HDx Includes "Excite HDi" system. 14.0 Software only.
- Software
- 14.0 For both Excite HDx and 3T Excite HDx systems.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.