Pular para o conteúdo principal
GE Healthcare · MRSigna2225518

Cannot Scan With TR and Dynamic Disable Faults.

Symptom

Cannot Scan With intermittent TR and Dynamic Disable Faults. The customer when scanning is getting intermittent TR and Dynamic Disable faults. It happens with all coils and cannot be pinned down to a specific PSD.

Error message

2225518 Dynamic disable and T/R driver fault. Possibly a shorted diode in the head T/R switch network, or a short circuit load. See long description Thu Feb 6 04:02:38 1997 2225518 1 lx-bay14b NSP HPC_PORT_ID:spi_tr_mon 6060 Dynamic disable and T/R driver fault. Possibly a shorted diode in the head T/R switch network, or a short circuit load.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR