Symptom
MOD Save or Restore functions fail. MOD Save or Restore functions fail Troubleshooting There is no media in the drive message appears in a attention box during a MOD Save or Restore of with Error: 10191 Failed save exam. posted in the Message Log. The fllowing pop-up message appears on screen. Problem is consistent If a different disk is used then the error occurs again when trying to save, restore, or label
Error message
Error: 10191 Failed save exam. Mon Nov 27 03:45:02 2006 Error: 10191 Host: emc1 Process: arserver File: arscommand.c Line: 2304 Failed save exam. Paused 16896 Error: 10033 There is no media in the drive. See Long Description for complete error Sun Aug 14 11:40:26 2005 Error: 10033 Host: ct01 Process:arsjob File: ArchiveDevice.c Line:4781 There is no media in the drive.
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (60kW) LightSpeed 5.X (60kW) ~ Use with MDAS or GDAS HP_P_G4 HP_P_G8 HP_P_G16 HP_P_M16 LightSpeed 5.X 60KW G-DAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.