GE Healthcare · MRSigna2244591
CWA ID81904. Will Not Scan with Gradient Errors.
Symptom
Will Not Scan with Gradient Errors.
Error message
2244591 SGA Power Supply Over Voltage Fault See Long Description. Fri May 7 16:14:56 2004 Error: 2244591 Host: NSM6 Process: NSP File: SCP:GP Line: 1857 SGA Power Supply Over Voltage Fault. FOR FIELD ENGINEER: Look at SGA-PS front LEDs. If BUS OV LED lit, check PDU 420V output. If OV X,Y,Z LED lit, swap military connectors to isolate SGA vs. SGA-PS. Power cycle ACGD POWER breaker on PDU to recover.