GE Healthcare · MRSigna2244591

CWA ID81904. Will Not Scan with Gradient Errors.

Symptom

Will Not Scan with Gradient Errors.

Error message

2244591 SGA Power Supply Over Voltage Fault See Long Description. Fri May 7 16:14:56 2004 Error: 2244591 Host: NSM6 Process: NSP File: SCP:GP Line: 1857 SGA Power Supply Over Voltage Fault. FOR FIELD ENGINEER: Look at SGA-PS front LEDs. If BUS OV LED lit, check PDU 420V output. If OV X,Y,Z LED lit, swap military connectors to isolate SGA vs. SGA-PS. Power cycle ACGD POWER breaker on PDU to recover.

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Area
MR

Full cause & step-by-step resolution