Gantry display shows "err" and fails firmware flash download

Symptom

Gantry display shows ERR and fails firmware flash download. See long description

Error message

Error: 182826 Communication Error: Communication Packet with the GANTRY DISPLAY. Open Long Description for full error Wed Dec 22 20:49:26 2004 Error: 182826 Host: ETC Process: Gantry Display File: gd_tnc_driver.c Line: 477 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Soft Time: 20:48:40:485 Log Series: 31 Communication Error: Communication Packet with the GANTRY DISPLAY. Message Code: 2. (Data = 0x20, 0x0, 0x0, 0x0).

Error: 182865 STUCK BUTTON detected on Gantry Controls See the LD for the full error code Tue Sep 13 18:09:37 2005 Error: 182865 Host: ETC Process: Gantry Display File: gd tnc erlog.c Line: 782 Function: No System Function Reported Scan: 9160/5/0 Type: None/Unknown Exception Level: Pri/Soft Time: 18:09:37:415 Log Series: 928 STUCK BUTTON detected on Gantry Controls . Node Id: 4 , Button: ALIGNMENT LIGHT BUTTON. Board Number: 2256728 Serial Number: 18795.

Applies to

Models
LightSpeed
Product
LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
2256730-3 Display assembly used on 2.X/3X/4X5X/7X See long description for details. Part #: 2256730-3 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: H-POWER COMPATIBLE GANTRY DISPLAY FRU: Yes Status: Orderable Comments: FLS - H-POWER COMPATIBLE GANTRY U/O CT H POWER DISPLAY
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.