GE Healthcare · CTLightSpeedVCT260133240
LightSpeed 7.x (VCT, VCT Select, Pro32)
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 260133240 A2D Limit Test failed- See error variations See Long Description for full error Fri Jun 22 17:31:33 2007 Error: 260133240 Host: orp Process: orp_subsys File: ADControl.cxx Line: 376 Fri Jun 22 17:31:32 2007 Host : Collimator Ermes # : 260133240 Exception Class : Secondary Severity : Pri/Soft File : ADControl.cxx Line# : 376 Function : No System Function Reported Scan Type : None/Unknown Scan: 0/0/0 A2D Limit Test failed for the P5VA channel Value returned from A2D : 988 millavolts Channel upper limit : 5250 millavolts Channel lower limit : 4750 millavolts. Wed Mar 14 12:16:13 2007 Error: 260133240 Host: orp Process: orp_subsys File: ADControl.cxx Line: 286 Wed Mar 14 12:16:13 2007 Host : Collimator Ermes # : 260133240 A2D Limit Test failed for the P48V A channel Value returned from A2D : 1897 millavolts Channel upper limit : 1775 millavolts Channel lower limit : 1605 millavolts.
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
- Component
- 5130200 VCT DCB PWA See long description for details Part #: 5130200 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: VCT DCB PWA FRU: Yes Status: Orderable Parts Offered as: Current: 5130200 Harvested: Not Available Exchange: Not Available Comments: Added per VCT RSPL, ms13452
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required