GE Healthcare · CTLightSpeed200002352245175
81072 LightSpeed 5.X(100kW) GDAS16 with GRE
Symptom
Applications software fails to boot up as it fails to initialize tube cooling library
Error message
Error: 200002352 A HOST_SHUTDOWN was initiated by the death of scanmgr See Long Description for full error Tue Nov 23 17:26:47 2004 Error: 200002352 Host: srct Process: cupMonitor File: ssProcStop.c Line: 1065 A HOST_SHUTDOWN was initiated by the death of scanmgr With 06MW03.X software this is listed as: Fri Oct 6 19:02:31 2006 Error: 200002352 Host: dominct Process: cupMonitor File: ssProcStop.c Line: 766 A HOST_SHUTDOWN was initiatied by the unexpected termination of scanmgr. Error: 245175 Failed to initialize the tube-cooling library See long description for complete error Fri Error: 245175 Host: ct01 Process: scanmgr File: scanmgr.cxx Line: 499 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Failed to initialize the tube-cooling library .
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.