Replace HFA CONTROL BOARD ASSEMBLY.

Symptom

Current distortion faults with no other fault present. May happen on one, two or three gradient axes at once.

Error message

2245053 GRADIENT DRIVER TESTS - X, Y or Z Axis Fault Occurred GP Reported a X,Y or Z Axis Current Distortion Fault. Check LCoil calibration. 2245053 "GRADIENT DRIVER TESTS - %c Axis Fault Occurred.\n\ GP reported a %c Axis Current Distortion Fault.\n\ Check LCoil calibration. Swap cables between SGA\n\ & GP to isolate if SGA or GP is the source of\n\ problem.\n\ 2245053 "GRADIENT DRIVER TESTS - %c Axis Fault Occurred. GP reported a %c Axis Current Distortion Fault. Check LCoil calibration. Swap cables between SGA & GP to isolate if SGA or GP is the source of problem.

Applies to

Models
Excite, Signa
Product
Excite HDx Includes "Excite HDi" system. 14.0 Software only.
Software
14.0 For both Excite HDx and 3T Excite HDx systems.
Component
2396050-2 HFA CONTROL BOARD ASSEMBLY
Area
MR

Full cause & step-by-step resolution