Replace HFA CONTROL BOARD ASSEMBLY. 3T Excite HD 12.0. 2244578 The GP is reporting a Gradient Amplifier % Axis Current D…

Replace HFA CONTROL BOARD ASSEMBLY. 3T Excite HD 12.0. 2244578 The GP is reporting a Gradient Amplifier % Axis Current Distortion Fault

Symptom

Current distortion faults with no other fault present. May happen on one, two or three gradient axes at once.

Error message

2245053 GRADIENT DRIVER TESTS - X, Y or Z Axis Fault Occurred GP Reported a X,Y or Z Axis Current Distortion Fault. Check LCoil calibration. 2245053 "GRADIENT DRIVER TESTS - %c Axis Fault Occurred.\n\ GP reported a %c Axis Current Distortion Fault.\n\ Check LCoil calibration. Swap cables between SGA\n\ & GP to isolate if SGA or GP is the source of\n\ problem.\n\ 2245053 "GRADIENT DRIVER TESTS - %c Axis Fault Occurred. GP reported a %c Axis Current Distortion Fault. Check LCoil calibration. Swap cables between SGA & GP to isolate if SGA or GP is the source of problem.

Applies to

Models
Excite, Signa
Product
3T Excite HD 12.0 software only
Software
12.0 Excite HD and 3T Excite HD products only
Component
2396050-2 HFA CONTROL BOARD ASSEMBLY
Area
MR