The problem appeared with Cine mode, which collected data in groups and produced series with lots of images (>1200) The recons would fail when it tried to recon the scans at the high end of the data c…
Symptom
The problem appeared with Cine mode, which collected data in groups and produced series with lots of images (>1200) The recons would fail when it tried to recon the scans at the high end of the data collection. Images from the beginning of the data collection recon did not fail. If they scanned regular head, abdomen, etc protocols, recon did not fail The problem appeared with Cine mode, which collected data in groups and produced series with lots of images (>1200) The recons would fail when it tried to recon the scans at the high end of the data collection. Images from the beginning of the data collection recon did not fail. If they scanned regular head, abdomen, etc protocols, recon did not fail
Error message
Error: 290009007 Timeout waiting for image available. Number of retry so far: 1. IG node: 0. Sun May 14 14:05:42 2006 Error: 290009007 Host: darc Process: recon_control File: ImageSetRequest.cxx Line: 5208 Function: Image Generation : SBC Processing Job Type: RECON_JOB_PROS_HELICAL_RECON Scan: 378/2/2 Image: 378/2/39 Timeout waiting for image available. Number of retry so far: 1. IG node: 0. Error: 290009002 Error failed in opening file See long description for full error Mon Jul 9 17:34:45 2007 Error: 290009002 Host: darc Process: recon_control File: IsrUpdateCache.cxx Line: 153 Error failed in opening file. file name = /usr/g/queue/priority3_jobs/FCCC.3776.2.1177.1183997163.967569.pro_cine error msg = No such file or directory
Applies to
- Models
- LightSpeed
- Product
- LightSpeed RT4 WB O G4 G HPT WB O M4 G HPT LightSpeed RT4 with GRE and MDAS4 or GDAS4 gantry At GE Medical Systems, we’re pioneering exciting CT technology for Radiation Therapy planning. Introducing the newest addition to our LightSpeed® series – LightSpeed RT, the first multi-slice, wide-bore CT scanner with the largest field of view in the industry. With a generous 80 cm gantry opening, the LightSpeed RT bore is 10 cm larger than conventional scanners. This provides complete flexibility for patient positioning freedom – even for the most demanding oncology needs. And the 65 cm variable field of view delivers 15 additional centimeters of visual coverage, allowing clinicians to see a patient’s entire anatomy for precise radiation therapy planning. Position and precision Thin, multi-slice imaging is the extra bonus – enabling respiratory gating, providing precise target delineation and delivering higher resolution digitally reconstructed radiographs (DRRs) than ever before. All adding up to the ability for you to deliver what really matters most – better cancer care for your patients. Plus, the LightSpeed RT scanner is available with the Exact™ couch insert, enabling you to consistently replicate patient immobilization for treatment. With LightSpeed RT, the future of your Radiation Oncology department is wide open. Links to the Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required