GE Healthcare · CTLightSpeed200281000
Images failing to display in Image Works
Symptom
Images are showing up on the monitor with autoview but not in Image Works. Archiving images and networking of the images are failing.
Error message
Error: 200281000 Description='AiM error: requestedService='Ex_Ser_Path_Rqst' message='Browser cannot find exams or series or images matching e/s/i request 1103/601'' Code=3 See LD for complete error Sat Dec 2 11:03:46 2006 Error: 200281000 Host: mhvct1 Process: visScript File: VavBrowserProxy.cxx Line: 567 Error reported: Class=VavBrowserProxy Method=answerToRequestCallback Location=VavBrowserProxy.cxx, 567 Description='AiM error: requestedService='Ex_Ser_Path_Rqst' message=' Browser cannot find exams or series or images matching e/s/i request 1103/601'' Code=3
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 4.X - Octane sw_version: 405I.2_H4.0M5 LightSpeed 4.X - Octane Gantry Model # 2339985 Console Model #'s 2341923 and 2341923-2(D3D) check_config shows LightSpeed 16 / LightSpeed 16 Product: LightSpeed16 sw_version: 405I.2_H4.0M5 or newer machine_type: Octane host_id: 3232297858 tube_type: 15 Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Software
- 405I.2_H4.0M5 for LightSpeed 4.X (LightSpeed 16) with H16 Console (GOC1)(2341923, 2341923-2) - showprods returns 405I.2_H4.0M5 for LightSpeed 16 with GOC1 SGI Octane CPU 405I.2_H4.0M5 for LightSpeed 16 with GOC1 SGI Octane CPU Class C Class M Application Software Set CD Part No. Ver. CD Label CD Part No. Ver CD Label Set Part No. Apps. CD Part No Apps. CD Label showprods/ Service Desktop Intro. Date Comment 2395172 1104I.2_ H4.0M5 CLASSC LightSpeed MDAS 4/8/16/ HiSpeed QXi Class C Software 2395173 1204I.2_ H4.0M5 CLASSM LightSpeed/ HiSpeed QX/I Class M Software 2395176 2395170, 2395171, 2399867 LightSpeed 16 Applications Software (Disk 1, Disk 2 and Patch A) 405I.2_H4.0M5 FW47-03 LS2002M5. IRIX only. Class C/M Common w/ LS 2X, 3X, & 4X.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required