Will Not Scan with Gradient Errors.

Symptom

Will Not Scan with Gradient Errors.

Error message

2226629 Gradient Identification Problem The gradient subsystem could not identify itself as 8915 or 8905 gradients. 2226629 Gradient Identification Problem. The gradient subsystem could not identify itself as 8915 or 8905 gradients. Possible gradient amplifier problem, please run diagnostics.

2244598 The GP Reported more than one %c Axis SGA Internal Wiring/Power Supply Undervoltage Faults 2244598 "The GP reported more than one %c Axis SGA Internal\n\ Wiring/Power Supply Undervoltage Faults. Refer to prior\n\ fault message. Power cycle the ACGD POWER breaker on the\n\ PDU to recover from this condition."

2244590 SGA-PS OverCurrent Fault 2244590 "SGA-PS OverCurrent Fault.\n\ Look at SGA-PS OC front LEDs to determine which SGA axis.\n\ If primary OC, replace SGA-PS. If secondary OC, swap military\n\ connectors to isolate SGA vs. SGA-PS. Power cycle ACGD POWER\n\ PDU breaker to recover."

2244573 The GP is reporting %c Axis SGA Internal Wiring Fault and Internal Power Supply Undervoltage 2244573 "The GP is reporting %c Axis SGA Internal Wiring Fault and\n\ Internal Power Supply Undervoltage. Perform a TPS Reset.\n\ Else, replace the SGA."

Applies to

Models
Signa
Product
Signa LX Signa LX Product can be used with 8.x and 9.x systems.
Software
9.1 For Signa LX product only
Component
2334862-2 ACGD SGA PS ASM W/FRU PACKAGING See Long Description 2334862-2 Part Type: GP REPAIRABLES Part Weight: 2.18 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ACGD SGA PS ASM W/FRU PACKAGING FRU: Yes Status: Orderable This supply is for use with HFD, HFD-S, and ACGD Cabinets. ** WARNING - POSSIBLE PERSONAL INJURY. HEAVY OBJECT - THE POWER SUPPLY WITH CASE WEIGHS APPROXIMATELY 240 LBS (109 KG). Assistance to lift this product mechanical (Dollies) or Additional personnel] is required. 212002950 May 27, 2009
Area
MR

Full cause & step-by-step resolution