Blowers and Fans Not Running in MKS 35KW RF Amplifier Chassis due to the C Phase 8-Amp Fuse Inside MKS 35KW RF Amplifier…

Blowers and Fans Not Running in MKS 35KW RF Amplifier Chassis due to the C Phase 8-Amp Fuse Inside MKS 35KW RF Amplifier Chassis is Blown.

Symptom

Troubleshooting 3T MKS RF Amplifier PA Air Flow fault 41 Troubleshooting 3T RF Amplifier PA Air Flow fault 41

Error message

2254256 3T RF Amplifier PA Air Flow (fault 41) see long description Sun Mar 18 17:42:13 2007 Error: 2254256 Host: MR0C1 Process: NSP File: SCP:../rf_nb_contro Line: 1491 3T RF Amplifier PA Air Flow (fault 41). Possible Causes: dirty air filters, bad fuses, defective sensor board. Check that air filters are clean. Test fuses on the Power Supply deck. If the fault persists, replace the RF amplifier.-> NOTE: This error can be cause by one of the 3 "hidden fuses" opening. See SN 67103 for details.

Applies to

Models
Excite
Product
3T Excite HD 12.0 software only
Software
12.0 Excite HD and 3T Excite HD products only
Component
2336064-8 3T CORONA MKS RF AMPLIFIER FUSE KIT 3T CORONA RF AMPLIFIER FUSE KIT 2336064-8 3T CORONA RF AMPLIFIER FUSE KIT 2336064-8
Area
MR

Full cause & step-by-step resolution