GE Healthcare · CTHiSpeed1831212257550013
HiSpeed Advantage RP/S
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 183121 FIL/MA fault detected: Open filament circuit. See the long descripton for the full error code Tue Oct 11 13:12:27 2005 Error: 183121 Host: OBC Process: Generator Function: XRAY GENERATION : Filament Selection Scan: 22575/2/1 Type: Helical Exception Level: Pri/Most Severe FIL/MA fault detected: Open filament circuit. Desired focal spot = 0. (0 = large, 1 = small) <== Large Filament is open OR Fri Sep 22 07:03:11 2006 Error: 183121 Host: OBC Process: Generator File: gen_abort.c 1.31 Line: 341 Function: XRAY GENERATION : Filament Selection Scan: 50013/10/40 Type: Axial Exception Level: Pri/Most Severe Time: 07:03:11:045 Log Series: 393 FIL/MA fault detected: Open filament circuit. Desired focal spot = 1. (0 = large, 1 = small)<-------- Small
Applies to
- Models
- HiSpeed
- Product
- HiSpeed Advantage RP/S HISPEED ADVANTAGE RP SOFTWARE LFC TAPE Version: rp25.5 Created: Tue Jan 26 09:23:55 CST 1999 Copyright: (c) 1988, 1999 The General Electric Company HISPEED ADVANTAGE RP SOFTWARE LFC TAPE Version: rp25.5 Created: Tue Jan 26 09:23:55 CST 1999 Copyright: (c) 1988, 1999 The General Electric Company Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- D3101T / D3102T X-ray Tube Used In CT/i and HiSpeed Advantage Part #: D3102T Part Type: GP TUBES Part Weight: 0.00 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: TUBE HSA/CTI FRU: Yes Status: Orderable Comments: FLS - NEW 1994 TUBE CAT. FOR RELOAD REPLACES D1199ZR (RELOADED) .9 X .7FS, 1.2-1.2 FS HISPEED ADVANTAGE DIMS 37 X 27 X 28 (LWH) MAXIRAY EQUIVALENT IS D3109T TUBE FAN= 46-219465P1 AM SOURCE = GEMS-MM (MEXICO) *** FLS Tube Names Systems Compatible Heat Capacity Focal Spot Catalog # Model # Insert # Casing # Tube Unit # HSA(Zeus) HiSpeed CT/I, or HiSpeed Advantage 3.5MHu 0.9x0.7/1.2x1.2 D3102T MX165 46-309300G2 46-309500G2 46-309500G801 Tubes started shipping half machined, half cast casings in early 2005. This was stated to be a tempory change in process while a new vendor was being evaluated.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required