Pular para o conteúdo principal
GE Healthcare · MRExcite2249317108193495622493071081934830200002249339

TPS Reset fails With AP initialization error Auto configuration failed.

Symptom

Unable to Download or Reset the TPS MGD or ISE Unable to download TPS or ISE or MGD chassis

Error message

2249317 Communication problem between the APS, AGP, and SCP boards. MGD Resource Probe could not find the shared data structure in the name database. 1081934956 2 1 Wed Apr 14 09:29:16 2004 2249317 2 CARD diagCoord systemResourceApi.c 357 Task ID: 0x3f1fdf8 Time: 8585 Object: APS Communication problem between the APS, AGP, and SCP boards. MGD Resource Probe could not find the shared data structure in the name database. vxWorks errno = 0x590001.->

2249307 Communication problem between the APS, AGP, and SCP boards. Bridge Board mapping synchronization timeout occured. Error seen with MGD Diags: 1081934830 0 1 Wed Apr 14 09:27:10 2004 2249307 2 CARD diagCoord pearlMap.c 1806 Task ID: 0x0 Time: 0 Object: SCP Communication problem between the APS, AGP, and SCP boards. Bridge Board mapping synchronization timeout occured. Value 0xAAAA not found at address 0xC03010A8 after 20000 msec.->

2249339 CM init failed. smNameAdd failed. 1081934956 5 1 Wed Apr 14 09:29:16 2004 2249339 2 CARD diagCoord cm_mp.c 410 Task ID: 0x3a652d0 Time: 10639 Object: APS CM init failed. smNameAdd failed. errno = 0x590001.->

2249338 CM init failed. msgQSmCreate failed when creating VxMP mailbox. Error seen with 

Applies to

Models
Excite
Product
Excite HDx Includes "Excite HDi" system. 14.0 Software only.
Software
14.0 For both Excite HDx and 3T Excite HDx systems.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.