GE Healthcare · CTLightSpeed1832966500118325718763183109
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane
Symptom
Scanner Hardware Fault with high voltage spits. Scanner Hardware Fault with high voltage spits
Error message
Error: 183296 Monitoring: (Measured 245.5 mA vs Ref 240 mA) See Long Description for the full error Error: 183296 Suite: CT01 Host: OBC Process: Generator File: gen_tech.c 1.36 Method: No Method Line: 529 Function: No System Function Reported Scan: 65001/8/38 Type: Axial Exception Level: Sec/Soft Time: 03:11:20:495 Log Series: 49 Selected Technic: 100 kV, 240 mA, 0 % Mod, Backup Timer 4324. VarmA: (Index: Start 0, Skip 0, Last 0) Last Cmd 240 mA. Monitoring: (Measured 245.5 mA vs Ref 240 mA) . 5 Total Samples, Tube Spit Ctr 0. (Rotor 0x0000ae04, Gen 0x9c19b5ff, Mon 0x00020000 Error: 183257 KV board tube spit counter = xx See Long Description for complete error 183257 KV board tube spit counter = 1 Tue Nov 30 15:38:17 2004 Error: 183257 Suite: CT01 Host: OBC Process: Control File: gen_abort.c 1.33 Method: No Method Line: 716 Function: XRAY GENERATION : KV Loop Scan: 18763/2/0 Type: None/Unknown Exception Level: Pri/Soft Time: 15:38:16:115 Log Series: 387 KV board tube spit counter = 1 Error: 183109 KV fault detected: Anode overcurrent See Long Description for complete error Tue Nov 30 15:38:17 2004 Error: 183109 Suite: CT01 Host: OBC Process: Generator File: gen_abort.c 1.33 Metho
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required