Pular para o conteúdo principal
GE Healthcare · CTLightSpeedVCT260116313244634

LightSpeed 7.x (VCT, VCT Select, Pro32)

Symptom

Hardware reset fails with Laser indicator flashing on the Gantry Display board. Hardware reset fails with Laser indicator flashing on the Gantry Display board

Error message

Error: 260116313 Table Gantry Processor - Host (OC) LAN Communication Watchdog Timed out. see LD Tue May 16 22:03:29 2006 Error: 260116313 Host: tgp Process: tgp_subsys File: subsys_mgr.c Line: 349 Running: System Reset Tue May 16 21:37:53 2006 Host : Table Gantry Processor Ermes # : 260116313 Exception Class : Safe State Severity : Secondary File : subsys_mgr.c Line# : 349 Function : SYSTEM COMMUNICATIONS : Control Network Scan Type : None/Unknown Scan: 0/0/0 Table Gantry Processor - Host (OC) LAN Communication Watchdog Timed out.

Error: 244634 Subsystem State; DCB:up, CCB:up, TCB:up, JEDI:up, --- TGP/ORP:down, TGP/ORP:DOWN See Long Description Sun Apr 13 14:45:14 2008 Error: 244634 Host: lawct2 Process: fwmgr File: fwmgr_utils.c Line: 664 Function: No system function reported Scan Type: Not scanning Scan: 0/0/0 Scan Seq Id: 0 Subsystem State; DCB:up CCB:up TCB:up JEDI:up TGP/ORP:down .

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 7.x (VCT, VCT Select, Pro32) V40_H_V64_G_GTM V40_H_V64_G_GTL V40_H_V32_G_GTM V40_H_V32_G_GTL V40_H_H64_G_GTM V40_H_H64_G_GTL V40_H_HS64_G_G Product Name: LightSpeed 7.x (VCT, VCT Select, Pro32) Product photographs Method of remote or software identification - See attachment for a complete list of software commands and typical results. check_config swhwinfo “swhwinfo” will have the following new switches available with the VCT M4 and PF 1.5 M3 software release (software release deploying this “swhwinfo” capability is “06MW03.?”): · -c or -classc = Outputs the Class C version, if installed, on the system · -m or -classm = Outputs the Class M version, if installed, on the system · -o or -os = Outputs the Operating System version installed on the system · -da or -date = Outputs the date and time the software was installed on the system (LFC) · -h or –help = Outputs complete details about all the available switches associated with swhwinfo · swhwinfo –c –m –o -da List of Supported Console Types GOC5 Will have a VDARC and VIG GOC6 Will have a VDIG List of softwares the product can run 09MW08.10 09MW08.11 08MW33.2 07MW18.4 06MW03.4 05MW31.6 FAQ's Table weight limit - Up to 500 lb. (227 kg) with ± .25 mm positional accuracy Tube type - D3194T D3193T(green) Current production - M4 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system. Done acording to template
Component
U8006FF Fuse (1 amp) See Long Description for Details U8006FF Fuse (1 amp) Part #: U8006FF Part Type: GP CONSUMABLES Part Weight: 0.00 LBR Repairable: No Returnable: No Hazardous Material: N Description: FUSE FRU: Yes Status: Orderable Comments: Not Available NOTE: This fuse is included in fuse kit of the following Systems: LS 7.x Fuse Kit # 5125429 BrightSpeed Fuse Kit # 2385412-2
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues