Pular para o conteúdo principal
GE Healthcare · MRExciteSigna2249161

Cannot scan with TR and dynamic disable fault Dynamic disable fault Due to SSM High Voltage SupplysCannot Scan With TR a…

Cannot scan with TR and dynamic disable fault Dynamic disable fault Due to SSM High Voltage SupplysCannot Scan With TR and Dynamic Disable Faults.

Symptom

Cannot Scan With intermittent TR and Dynamic Disable Faults. The customer when scanning is getting intermittent TR and Dynamic Disable faults. It happens with all coils and cannot be pinned down to a specific PSD.

Error message

2249161 Dynamic disable fault Rear of Coil short 2249161 Host: pomr Process: NSP File: SCP:../TrDriverMdsHWO.cpp Line: 1699 Dynamic disable fault, Rear of Coil short^A Dynamic Disable bias fault on back drivers. Possibly a short circuit load or a shorted diode in the dynamic disable switch boxes on BACK of BRM.

Applies to

Models
Excite, Signa
Product
Excite Includes Excite 1 and Excite 2, 10.0, 11.0 and 11.1 Software.
Software
11.0 Only for 1.5T Excite product. 12.0 is Excite HD and 14.0 is Excite HDx. Note, 3T Excite goes with software G3, not 11.0 or 11.1.
Area
MR

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare MR issues