OrthoMRI Mechanical swing arm is broken
Symptom
OrthoMRI The coil ID is not working on all coils.
wqr
See Long Description OLEs/RSE's are required to gather Coil Serial number and Part number and place the info in the Tech log box for Diagnosis and Troubleshooting: If the information is available to the customer without affecting their current work flow, the OLE should attempt to gather this information when the call is not split (paged) to the FE . Otherwise the OLE should state that " the coil or system is in use at time of call and the coil information is not currently accessible ". If the RFS is properly split it should link the next RFS and the FE would be able to provide this information. FEs are required to gather Coil Serial number and Part number and answer the Coil question in FEMC debrief where they will place the info in the PopUp Box. Proprietary, For GE Healthcare personnel Only This information is Property of GE Healthcare - For GE Healthcare Service Personnel Only. Disclosure to Third Parties Prohibited.
wqr
212002465 6 May 2011
Error message
OrthoMRI Unknown Coil Error The system does not recognize the coil that is plugged in for use. The high failure part is the coil cable. If this error occurs on more than one coil, the mechanical swing arm cables is the next high failure part.
Applies to
- Product
- OrthoMRI EscanNE. OrthoMRI EscanNE.
- Software
- OrthoMRI S/W Release 9.1B. OrthoMRI S/W Release 9.1B.
- Component
- OrthoMRI LNR9101821001 Moving mechanism swing arm with cables
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.