GE Healthcare · MR
OrthoMRI Room temperature is 85 degrees
Symptom
OrthoMRI Getting signal saturation error
Error message
OrthoMRI 1-46 Signal Saturation Do you want to proceed? Try to perform these steps in this order: check the Magnet Temperature, check Magnet Heaters and Sensors resistor values, change TMP and or MTCM. Typical issue is the software not at 9.1b or higher revision or the patient's skin is touching the coil. Foam padding is needed to isolate the patient from the coil.
Applies to
- Product
- OrthoMRI CscanNE. OrthoMRI CscanNE.
- Software
- OrthoMRI 8.2b software Revision. OrthoMRI 8.2b software Revision.
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.