GE Healthcare · CTLightSpeed183365501681832621831825000144858
LightSpeed 2.X (Plus, QX/i) - Octane
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 183365 Cannot bring HVDC bus up for xx seconds See long Description for complete error Wed Apr 5 06:52:02 2006 Error: 183365 Host: OBC Process: Generator File: gen_rotor.c 1.39 Line: 320 Function: XRAY GENERATION : Tube Rotor Control Scan: 50168/5/8 Type: Static Exception Level: Pri/Most Severe Time: 06:52:02:815 Log Series: 118 Cannot bring HVDC bus up for 6 seconds. Error: 183262 Backup contactor opened without firmware command See the long description for the full error Mon May 17 12:20:18 1999 Error: 183262 Suite: CT01 Host: OBC Process: Generator File: ctvrc_manager.c 1.8 Method: No Method Line:823 Function: XRAY GENERATION : 550 VDC Backup Contactor Intlk Scan: 7073/3/15 Type: Axial Exception Level: Pri/Most Severe Time: 12:20:18:495 Log Series:513 Backup contactor opened without firmware command. Error: 183182 I/O board pilot relays are closed. PDU pilot relay not energized. HVDC Rail Voltage is xxxx See Long Description for more details. Error: 183182 I/O Bd. pilot relays are closed. PDU pilot relay not energized. HVDC is xxxx. ** Note the HVDC Rail Voltage. It is important in troubleshooting Tue Sep 21 06:13:59 2004 Error: 183182 Host: OBC Process: Generator Funct
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 2.X (Plus, QX/i) - Octane Always S-DAS 2.4.2_H2.4M5 LightSpeed 2.X (Plus, QX/i) - Octane (Pegasus or Rhapsode IG) - Always S-DAS Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 2.4.2_H2.4M5 FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 FEATURES/OPTIONS STD/OPTION Cat # Digital Laser Camera I/F B7700PD Smart Helical (s/w only) Standard B7530TK Fast Recon (s/w only) Option B7700JF BMD (requires B7710MP or MN) Option B7710MK Smart Prep Option B7530SP Smart Score ? ? 3D Analysis (Requires D3D) Option B7540TD Navigator (s/w only) Option B7540N DentaScan (s/w only) Option B7540SD Direct 3D (Requires B7710SA) Option B7810SS HIS/RIS Option (h/w and s/w) Option B7500PN Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-296221P15 80A 600V Fuse See Long Description for details Part #: 46-296221P15 Part Type: GP CONSUMABLES Part Weight: 0.13 LBR Repairable: No Returnable: No Hazardous Material: N Description: BUSSMANN KBC80 FRU: Yes Status: Orderable Comments: FLS - CLASS 1 FRU. PDV BUSSMANN KBC80 80A, 600V THERE ARE 3 OF THESE FUSES IN FUSE KIT 2218570 - MS4751 8/01 PRICE CHANGE PER BRIAN NAGLE 4/98 FLS - FE ASSIST: IN ORDER TO INCREAS INCREASE REVENUE GLOBAL PARTS HAS CREATED A PROGRAM TO HARVEST PARTS FROM RENEWABLE RESOURCRESOURCES. THESE PARTS OFFER DIRECT SAVINGS TO THE BUSINESSAND THE FIELD. ORDER THE U PART WHEN AVAILABLE. IF NO U USED PARTS ARE AVAILABLE CARES WILL CHANGE THE ORDER TOA NEW PART.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required