GE Healthcare · CTHiSpeed183262

Chattering noises heard inside the PDU and K2 contactor chatters every time it is energized / Clean or replace as needed the contacts the PDU K2 auxillary contact sense switch.

Symptom

Chattering noises heard inside the PDU and K2 contactor chatters every time it is energized

Error message

Error: 183262 Backup contactor opened without firmware command See the long description for the full error Mon May 17 12:20:18 1999 Error: 183262 Suite: CT01 Host: OBC Process: Generator File: ctvrc_manager.c 1.8 Method: No Method Line:823 Function: XRAY GENERATION : 550 VDC Backup Contactor Intlk Scan: 7073/3/15 Type: Axial Exception Level: Pri/Most Severe Time: 12:20:18:495 Log Series:513 Backup contactor opened without firmware command.

Applies to

Models
HiSpeed
Product
HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Component
46-296690P1 auxillary contact sense switch on the side of PDU K2 contactor 46-186852P2 See long description for details Part #: 46-296690P1 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: PIGGY SWITCH K2 RELAY FRU: Yes Status: Orderable Comments: FLS - N.O. AUXILLIARY CONTACT (INTERLOCK) USED ON ALL HS PDU'S FRU CHANGE PER DUANE D. 10/3/97 PCN TO FOLLOW 2WENN 10/3/97 MS08992 PCN 199098 MS4636 10/20/97 ************************************************************CPD USE ONLY: IF NO PARTS AVAILABLE, BROKEN SWITCH CAN BE REPAIRED. FE # 1065 FOR INSTRUCTIONS IF INTERESTED- 12/28/98 MS13668******************************************** AVAIL. FROM VC536366 VNDR P/N-476-00004 JM/A3040 TRY ADVACOM AS POSSIBLE VENDORLAST PO 1995 OEM JOSLYN CLARK CONTROLS #5M63
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required