Chattering noises heard inside the PDU and K2 contactor chatters every time it is energized / Clean or replace as needed the contacts the PDU K2 auxillary contact sense switch.
Symptom
Chattering noises heard inside the PDU and K2 contactor chatters every time it is energized
Error message
Error: 183262 Backup contactor opened without firmware command See the long description for the full error Mon May 17 12:20:18 1999 Error: 183262 Suite: CT01 Host: OBC Process: Generator File: ctvrc_manager.c 1.8 Method: No Method Line:823 Function: XRAY GENERATION : 550 VDC Backup Contactor Intlk Scan: 7073/3/15 Type: Axial Exception Level: Pri/Most Severe Time: 12:20:18:495 Log Series:513 Backup contactor opened without firmware command.
Applies to
- Models
- HiSpeed
- Product
- HiSpeed CTi 6.X HiSpeed CTi 6.X with Octane and HSA Gantry HiSpeed CTi 6.X with Octane and HSA Gantry Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-296690P1 auxillary contact sense switch on the side of PDU K2 contactor 46-186852P2 See long description for details Part #: 46-296690P1 Part Type: GP CONSUMABLES Part Weight: 0.45 LBR Repairable: No Returnable: No Hazardous Material: N Description: PIGGY SWITCH K2 RELAY FRU: Yes Status: Orderable Comments: FLS - N.O. AUXILLIARY CONTACT (INTERLOCK) USED ON ALL HS PDU'S FRU CHANGE PER DUANE D. 10/3/97 PCN TO FOLLOW 2WENN 10/3/97 MS08992 PCN 199098 MS4636 10/20/97 ************************************************************CPD USE ONLY: IF NO PARTS AVAILABLE, BROKEN SWITCH CAN BE REPAIRED. FE # 1065 FOR INSTRUCTIONS IF INTERESTED- 12/28/98 MS13668******************************************** AVAIL. FROM VC536366 VNDR P/N-476-00004 JM/A3040 TRY ADVACOM AS POSSIBLE VENDORLAST PO 1995 OEM JOSLYN CLARK CONTROLS #5M63
- Area
- CT
Full cause & step-by-step resolution