GE Healthcare · MRExcite2252305
Loose head heliax connection at the penetration panel.
Symptom
Cannot scan in Head Transmit Mode with Head TR Driver Faults. Body Transmit works OK.
Error message
2252305 TR Driver Fault Detected Open Circuit. Source of Fault is Head TR 2252305 lx-mr NSP SCP:DCB 1240 TR Driver Fault Detected - Open Circuit. Source of Fault is Head TR. Value: 0 mA Limit: 900 mA->
Applies to
- Models
- Excite
- Product
- 3T Excite HD 12.0 software only
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.