GE Healthcare · CTHiSpeed18340695121603831184162
Scanner Hardware Stopped Scan
Symptom
Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan
Error message
Error: 183406 Filter home position exceeded MAXIMUM error tolerance See LD for full Error Error: 183406 Filter home position exceeded MAXIMUM error tolerance Mon Sep 13 10:28:02 2004 Error: 183406 Suite: CT01 Host: OBC Process: Collimator File: home_filter.c 1.15 3/3/99 Method: No Method Line: 131 Function: XRAY GENERATION : Xray Collimation and Filtration Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 10:27:43:860 Log Series: 26 Filter home position exceeded MAXIMUM error tolerance 951216 0446 03831 Current position = 164 encoder cnts. Desired position = 0 encoder cnts. Tolerance = 10 encoder cnts Error: 184162 Axial: Ok Collimator: Timeout Table: Ok Generator: Ok SARQ: Ok See LD for full error Error: 184162 Axial: Ok Collimator: Timeout Table: Ok Generator: Ok SARQ: Ok Sat Dec 16 04:33:57 1995 Error: 184162 Suite: GECT Host: STC Proc: Scan Control File: scReRefMgr.c 1.22 Method: No Method Line: 222 951216 0446 03831 Function: No System Function Reported Scan: 3102/1/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 04:33:56:985 Log Series: 172 Error detected in end-of-series operations. HW/SW Error: Possible hardware (reference) or software error
Applies to
- Models
- HiSpeed
- Product
- HiSpeed CT/i 4.X HiSpeed CT/i 4.x with Zeus HSA Gantry and SGI Indigo Console HiSpeed CT/i with:HSA GANTRY CT/I (RHAPSODE) CONSOLE (Indigo Computer) Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-296300G4 and 46-296300G5 EMC COLLIMATOR -used on CTi and compatible to HSA See LD for more info 46-296300G4 and 46-296300G5 EMC COLLIMATOR -used on CTi and compatible to HSA picture 46-296300G4 EMC COLLIMATOR Part #: 46-296300G4 Part Type: GP REPAIRABLES Part Weight: 19.54 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: 860-950117 EMC COLLIMATOR FRU: Yes Status: Orderable Comments: FLS - FE: READ SERVICE NOTE T-1379 5/96 GVR T3202GR MPS INFO: B7999KK CONTAINS G4 COLLIMATOR PLUS S/W. MFG PLANIS TO DEINSTALL THE G3, HAVE FE RETURN, UPGRADE, AND USE INADDITIONAL KITS PER R. KOPP. 10/95 FLS - FE ASSIST MS6588 5/2000 46-296300G4 = GRAPHITE FILTER COLLIMATOR. IF NO STOCK IS AVAILABLE ORDER 46-296300G5. THE G5 COLLIMATOR IS COMPATIBLE ON CTI HSARP AND HSAZ SERIES SYSTEMS. DO YOU STILL WANT TO ORDER THIS PART??? Part #: 46-296300G5 Part Type: GP REPAIRABLES Part Weight: 19.09 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: EMC COLLIMATOR FRU: Yes Status: Orderable Comments: FLS - COST REDUCED EMC COLLIMATOR USED ON FAST RHAPSODE 7/2/97 GWENN 08992 T3202KM 11/19/02 -BUILD OUTSOURCED FRMCT MFG TO ERDA PER GINA DUNDUNM KRAETSCH. MS12464 FLS - FE ASSIST MS6588 5/2000 T3202KM = G5 COLLIMATOR COST REDUCED EMC COLLIMATOR USED ONHSARP, HSA AND CTI SYSTEMS. THE G5 (T3202KM) WILL NOT WORKON HELIOS/LIGHTSPEED SCANNERS. DO YOU STILL WANT TO ORDER THIS PART???
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required