GE Healthcare · CTLightSpeed18317923000652925088230006530000000800000
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane
Symptom
Multiple recon retries and suspended images during an exam. See long description for details Multiple recon retries and suspended images during an exam. To get the images to recon you would have to resume the queue or reboot the system for the images to recon.
Error message
Error: 183179 I/O board sensed external x-ray interlocks not satisfied. Sun Sep 5 11:10:28 2004 Error: 183179 Host: OBC Process: Generator File: gen_control.c 1.39 Line: 658 Function: XRAY GENERATION : Xray Exposure Interlocks Scan: 4167/1/1 Type: Scout Exception Level: Pri/Soft Time: 11:10:29:165 Log Series: 41 I/O board sensed external x-ray interlocks not satisfied. Possible causes: - Scanning room door is open (if applicable). - Check relay on DIP and/or Axial board. This error is flagged by the OBC on I/O board during xray generation. Error: 230006529 Recon Mgr is sending queue ack to queuemgr with a RECON_JOB_FAILED_RETRY status See Long Description for Full Error Tue Feb 1 06:33:34 2005 Error: 230006529 Host: ice Process: reconmgr File: rm_recon_events.c Line: 329 Function: Image Generation : SBC Processing Job Type: Prospective Axial Scan: 25088/2/9 Image: 25088/3/25 Recon Seq Id: 957 Recon Mgr is sending queue ack to queuemgr with a RECON_JOB_FAILED_RETRY status Error: 230006530 RECON_JOB_FAILED_RETRY See Long Description for complete error Tue Feb 1 06:33:34 2005 Error: 230006530 Host: ice Process: reconmgr File: rm_recon_mgr.c Line: 1114 Function: Image Generation : SB
Applies to
- Models
- LightSpeed
- Product
- LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Software
- 308I.2_H3.1M5 for LightSpeed 3.X (Ultra, Plus, QX/I with H3 Console Assembly (2266832-3, 2304732-3) or GOC1-Octane2 Console (2341104, -2) - showprods returns LightSpeed Apps ver 308I.2_H3.1M5 308I.2_H3.1M5 for LightSpeed 3.X (Ultra, Plus, QX/I with H3 Console Assembly (2266832-3, 2304732-3) or GOC1-Octane2 Console (2341104, -2) - showprods returns LightSpeed Apps ver 308I.2_H3.1M5 308I.2_H3.1M5 for LightSpeed 3.X (Ultra, Plus, QX/I with H3 Console Assembly (2266832-3, 2304732-3) or GOC1-Octane2 Console (2341104, -2) - showprods returns LightSpeed Apps ver 308I.2_H3.1M5 Class C Application Software Set CD Part No. Ver. CD Label CD Part No. Ver. CD Label Set Part No. Apps. CD Part No Apps. CD Label showprods/ Service Desktop Intro. Date Comment 2395172 1104I.2_ H4.0M5 CLASSC LightSpeed MDAS 4/8/16/ HiSpeed QXi Class C Software 2395177 2395162, 2395163, 2399867 LightSpeed Plus/ Ultra MDAS Apps Software (Disk 1, Disk 2 and Patch A) 308I.2_H3.1M5 FW47-03 LS2002M5. IRIX Only. Class C/M Common w/ LS 2X, 3X, & 4X.
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.