Pular para o conteúdo principal
GE Healthcare · CT182125159854615983501232092312320727182105

Lightspeed 1.X - Octane

Symptom

Scanner Hardware Stopped Scan. Scanner Hardware Stopped Scan

Error message

Error: 182125 LAN Watchdog TIMEOUT with OBC or ETC (uuid 93, 193) (SW opening abort relay). See LD for full errors Mon Dec 6 22:19:33 2004 Error: 182125 Suite: CT01 Host: STC Process: Master Exc Handler File: exm_lan_wdog.c Method: No Method Line: 773 Function: No System Function Reported Scan: 0/0/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 22:19:32:730 Log Series: 27 LAN Watchdog TIMEOUT with OBC (uuid 193) (SW opening abort relay). HW/SW Error: Did not receive LAN Watchdog message from OBC. (Last Wdog Probe Sent:1598546 Last Response Rcvd:1598350 (in Ticks)). Error could be for the ETC also Reference Thu Jan 12 09:54:50 2006 Error: 182125 Host: STC Process: Master Exc Handler File: exm_lan_wdog.c Line: 880 Function: No System Function Reported Scan: 440/4/0 Type: None/Unknown Exception Level: Pri/Most Severe Time: 09:54:50:615 Log Series: 247 LAN Watchdog TIMEOUT for uuid 93 (193=OBC; 93=ETC). HW/SW Error: Did not receive LAN Watchdog message. (Last Wdog Probe Sent:12320923 Last Response Rcvd:12320727 (in Ticks)).

Error: 182105 LAN Watchdog TIMEOUT for uuid 93 (193=OBC; 93=ETC). See LD for error example. Tue Oct 25 22:34:14 2005 182105 3 CT01 STC Master Exc Hand

Applies to

Product
Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.