GE Healthcare · MRExcite2244591
Will Not Scan with Gradient Errors.
Symptom
Will Not Scan with Gradient Errors.
Error message
2244591 SGA Power Supply Over Voltage Fault See Long Description. Fri May 7 16:14:56 2004 Error: 2244591 Host: NSM6 Process: NSP File: SCP:GP Line: 1857 SGA Power Supply Over Voltage Fault. FOR FIELD ENGINEER: Look at SGA-PS front LEDs. If BUS OV LED lit, check PDU 420V output. If OV X,Y,Z LED lit, swap military connectors to isolate SGA vs. SGA-PS. Power cycle ACGD POWER breaker on PDU to recover.
Applies to
- Models
- Excite
- Product
- Excite HD For 12.0 1.5T Systems only.
- Software
- 12.0 Excite HD and 3T Excite HD products only
- Component
- 2334862-2 ACGD SGA PS ASM W/FRU PACKAGING See Long Description 2334862-2 Part Type: GP REPAIRABLES Part Weight: 2.18 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: ACGD SGA PS ASM W/FRU PACKAGING FRU: Yes Status: Orderable This supply is for use with HFD, HFD-S, and ACGD Cabinets. ** WARNING - POSSIBLE PERSONAL INJURY. HEAVY OBJECT - THE POWER SUPPLY WITH CASE WEIGHS APPROXIMATELY 240 LBS (109 KG). Assistance to lift this product mechanical (Dollies) or Additional personnel] is required. 212002950 May 27, 2009
- Area
- MR
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare MR issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare MR issues
- CWA How to Change Date or Time on LX, Excite, HD, HDe, HDx, HDxT, HFO, Excite HFO, Ovation1-4, Profile1-5, Discovery MR450 MR750 and Optima 450w Systems.
- CWA No Power to the Table and Magnet Display, Table Will Not Move And No Lights On the Magnet.
- CWA How to Change Exam Number from Remote Tools.
- Known Software issue with 15M4B coding error.
- Unable to Send Images to the PACS Due to PACS is down.
- U41 of the Slave RF Ctrl Assy, part of the PSU in the SRFD2 RF Amplifier Failed.