Revolution XRd and XQi Workstation Application won't launch - Bar graph keeps cycling 87% to 25%
Symptom
XRd and XQi When the Workstation application won't launch and the bar graph stops at 87%, check the EthernetDriverFile.cfg before reloading software. In several cases, this file was found to be missing entirely or corrupt (file name existed but content was 0 instead of 706 bytes).
Applies to
- Models
- Revolution
- Product
- Revolution XRd Revolution XQi
- Component
- ADS
- Area
- X-ray R&F
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray R&F issues
- ADx-E Advantx-E console displayed a 6C00 Error (Digital Imaging Subsystem error - Error status received in IDC Prep Reply) when the first exposure of detector bad pixel calibration was attempted; cali…
- Revolution XR/d After a Load From Cold, the AWS exhibits strange operation and the SUIF does not have any Positioner Configuration and Calibration selections due to the use of the incorrect Jumpstart…
- Revolution XRd2X Worklist query on "All Systems in this Modality" does not retrieve anything after FMI 10791A is completed due to the default query set to DX.
- Revolution XR/d error 4280 when trying to make exposure to table or wallstand detector
- Revolution XRd 2X The HP Workstation reboots itself.
- Def8000, Def5000, Revolution XRd and XRD2X Latent lead marker image burned into the detector.