GE Healthcare · CTHiSpeed183610183621
Scanner hardware stopped DAS PPI link ier error Unable place DAS in external/run mode / Replace EDAS Processor Board.
Symptom
Scanner hardware stopped scan with DAS PPI link error detected Scanner hardware stopped scan appears on the monitor WITH ERROR: 183621 DAS PPI link ier error 402 detected & Error: 183610 Unable place DAS in external/run mode
Error message
Error: 183610 Unable place DAS in external/run mode Mon Sep 19 04:19:05 2005 Error: 183610 Suite: SRC2 Host: OBC Process: DAS File: Method: No Method Line: 503 Function: SYSTEM COMMUNICATIONS : DAS Serial Link Scan: 0/0/1 Type: Static Exception Level: Pri/Most Severe Time: 04:19:05:670 Log Series: 41 Unable place DAS in external/run mode. Function call return status = 402. Error: 183621 DAS PPI link ier error 402 detected Mon Sep 19 04:19:05 2005 Error: 183621 Suite: SRC2 Host: OBC Process: DAS File: Method: No Method Line: 463 Function: SYSTEM COMMUNICATIONS : DAS Serial Link Scan: 0/0/1 Type: Static Exception Level: Pri/Soft Time: 04:19:05:665 Log Series: 40 DAS PPI link ier error 402 detected Application will retry communication attempt. Retry Count = 2 out of 2 possible.
Applies to
- Models
- HiSpeed
- Product
- HSA RP 2.X HiSpeed Advantage HiSpeed Advantage RP 2.X HiSpeed Advantage RP 2.X FMI 25247 HSA/RP 2.5 Y2K Upgrade Special Tools Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 46-183027P91 EDAS Processor Board - used on Full Gain and Half Gain EDAS See LD for FEMC data and picture EDAS Processor Board - used on Full Gain and Half Gain EDAS picture below Part #: 46-183027P91 Part Type: GP REPAIRABLES Part Weight: 0.45 LBR Repairable: Yes Returnable: Yes Hazardous Material: N Description: E-DAS PROCESSOR CARD FRU: Yes Status: Orderable Comments: FLS - PER M. GRABOWSKI TOP850 PROG. CHGD 1/95 KJK
- Area
- CT
Full cause & step-by-step resolution