Pular para o conteúdo principal
GE Healthcare · CTLightSpeed21408024294183182183179

LightSpeed 3.X (Ultra, Plus, QXi) - Octane

Symptom

Scan abort during any scan with external x-ray interlocks not satisfied.

Error message

Error: 214080 HVDC Bus voltage is present while Backup Contactor pilot relay indicates a de-energized state. Gentry I/O relays K251 and K245 are energized . HVDC Bus voltage : 35 Mon Mar 27 09:40:09 2006 Host : OBC Error: 214080 File: OBC diag HW Checks.c 1.36 Line: 297 Function: Diagnostics XRAY GENERATION : HVDC Backup Contactor Intlk Test Name: HV Backup Times UniqueID: 738 Iteration : 1 Exec Level: Hard Time : 24294 Error Type: Hardware HVDC Bus voltage is present while Backup Contactor pilot relay indicates a de-energized state. Gentry I/O relays K251 and K245 are energized . HVDC Bus voltage : 35 Genty I/O address FFAFBFH Bit: D4 Expected : 1 Actual : 0 May also see the following error in gesyslog: Mon Aug 16 14:39:13 2004 Error: 183182 Host: OBC Process: Generator File: gen rotor.c 1.39 Line: 498 Function: SYSTEM POWER CONTROL : HVDC Control Scan: 1935/1/1 Type: Axial Exception Level: Pri/Most Severe Time: 14:39:13:450 Log Series: 45 I/O board pilot relays are closed. PDU pilot relay not energized. BCSTAT = 0 HVDC Rail voltage = 195V Gentry I/O address: FFAFBFH Bit: D4 Expected: 1 Actual: 0 This error is flagged by the OBC Process: Generator on the HVDC Control during High V

Applies to

Models
LightSpeed
Product
LightSpeed 3.X (Ultra, Plus, QX/i) - Octane Always 4 or 8 slice MDAS 308i.2_H3.1M5 LightSpeed 3.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types Octane with Pegasus Recon hardware. List of softwares the product can run 308I.2_H3.1M5 305.4_H3.1M4 - obsolete 303.1_H3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.

7-day free trial · no credit card required

Related GE Healthcare CT issues