GE Healthcare · X-ray R&FRevolution2254537
XQ/i Revolution System uses Part Number 2254537 D-Link DES-1016 Ethernet Switch with a fiber optic plug-in mod…
Revolution XQi Ethernet Switch Part Issue, need to order fiber optic converter
Symptom
XQ/i Revolution System uses Part Number 2254537 D-Link DES-1016 Ethernet Switch with a fiber optic plug-in module in the Systems Cabinet - See attached picture (1): When FEs order this part, they receive a D-Link DSS-16+ Ethernet Switch. This is not a direct replacement; it is a different physical size and does not have a fiber optic interface. See attached picture (2).
Applies to
- Models
- Revolution
- Product
- Revolution XQi
- Area
- X-ray R&F
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray R&F issues
- ADx-E Advantx-E console displayed a 6C00 Error (Digital Imaging Subsystem error - Error status received in IDC Prep Reply) when the first exposure of detector bad pixel calibration was attempted; cali…
- Revolution XR/d After a Load From Cold, the AWS exhibits strange operation and the SUIF does not have any Positioner Configuration and Calibration selections due to the use of the incorrect Jumpstart…
- Revolution XRd2X Worklist query on "All Systems in this Modality" does not retrieve anything after FMI 10791A is completed due to the default query set to DX.
- Revolution XR/d error 4280 when trying to make exposure to table or wallstand detector
- Revolution XRd 2X The HP Workstation reboots itself.
- Def8000, Def5000, Revolution XRd and XRD2X Latent lead marker image burned into the detector.