XRD2X Very noisy tube rotor.
Revolution XRd2X (Lightning) EDI Error 3601200
Symptom
XRD2X Very noisy tube rotor. Replaced X-Ray tube, restored database; started to calibrate the tube; received system Error 3601200 (60-0325 DSP Overload). Small filament in the tube was open in the new tube. Replaced tube and received Error 3601200. Replaced kV Control Board (Lightning Jedi) with no difference.DSP error indicates that the Inverter frequency is at maximum. Components involved are the Main PS, Inverter, the HV transformer, Gate Cmd Board.
Error message
Error 3601200 Generator error 60-0325; (3): DSP overload
Applies to
- Models
- Revolution
- Product
- Revolution XRd2X
- Component
- JEDI Generator This generator is used on Proteus and Precision 500D systems.
- Area
- X-ray R&F
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray R&F issues
- ADx-E Advantx-E console displayed a 6C00 Error (Digital Imaging Subsystem error - Error status received in IDC Prep Reply) when the first exposure of detector bad pixel calibration was attempted; cali…
- Revolution XR/d After a Load From Cold, the AWS exhibits strange operation and the SUIF does not have any Positioner Configuration and Calibration selections due to the use of the incorrect Jumpstart…
- Revolution XRd2X Worklist query on "All Systems in this Modality" does not retrieve anything after FMI 10791A is completed due to the default query set to DX.
- Revolution XR/d error 4280 when trying to make exposure to table or wallstand detector
- Revolution XRd 2X The HP Workstation reboots itself.
- Def8000, Def5000, Revolution XRd and XRD2X Latent lead marker image burned into the detector.