Octane would not boot and monitor screens turn blank with PANIC
Symptom
System would not boot with PANIC: KERNEL FAULT reported. System would not boot and monitor screens turn blank with PANIC: KERNEL FAULT.
Error message
Memory Parity Error in SIMM Sxx Recoverable memory parity error detected by CPU PANIC: KERNEL FAULT - found in SYSLOG See LD for more info Memory Parity Error in SIMM S10 Recoverable memory parity error detected by CPU PANIC: KERNEL FAULT Defective SIMM memory on CPU board caused Memory Parity Error in the SYSLOG Troubleshooting kaiserct 12# cd /usr/adm kaiserct 13# tail -100 SYSLOG | more Jul 11 12:35:05 3D:kaiserct ppp[308]: + killall -v -USR1 gated Jul 11 12:35:05 3D:kaiserct ppp[308]: + 1>& 2 Jul 11 12:35:05 3D:kaiserct ppp[308]: UX:killall: ERROR: process <gated> not found Jul 11 12:35:06 2E:kaiserct savecore: reboot after panic: <0> PANIC: KERNEL FAULT Jul 11 12:35:06 2E:kaiserct savecore: pb 0: 8 <0x380> code:30 Jul 11 12:35:06 2E:kaiserct savecore: pb 1: Memory Parity Error in SIMM S10 Jul 11 12:35:06 2E:kaiserct savecore: pb 2: Recoverable memory parity error detected by CPU Jul 11 12:35:06 2E:kaiserct savecore: pb 3: Memory Parity Error in SIMM S10 TYPE THE FOLLOWING FOR SUSPECTED “MEMORY ERRORS ONLY” IN THE SYSLOG kaiserct 12# cd /usr/adm kaiserct 11# cat SYSLOG | grep Memory Jul 11 12:12:17 2A:kaiserct unix: Memory Parity Error in SIMM S10 Jul 11 12:13:26 2A:kaiserct un
Applies to
- Product
- Lightspeed 1.X - Octane 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete Lightspeed 1.X - Octane Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config showprods|grep Apps note: if check_config does not work then the software is not up to date with all FMI's. Check FMI history and verify that all patches have been loaded since last software reload. Ref FMI 25269 List of Supported Console Types Octane with Pegasus Recon hardware. may have an 02 SBC - Run df -k to see if the sbc is partitioned. List of softwares the product can run 10.5_2.8.2I_H1.3M4 10.4.1_2.8.2I_H1.3M4 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3176T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Example: English versions 2211223-100 Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2199806-8 Memory, 2x256MB Memory, 2x256MB part# 2199806-8 Memory, 2x256MB part# 2199806-8
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required