P500D Question about CAN termination jumpers for systems without an OTS or Wallstand?
Precision 500D Corona2 board does not need 120 ohm Can Bus termination
Symptom
P500D Question about CAN termination jumpers for systems without an OTS or Wallstand? There should never be any CAN termination jumpers installed for the Corona 2 with or without OTS and Wallstand. Odetta and Lightning required CAN termination jumpers, but Corona 2 based Thunder and Aurora systems do not. In fact, it was advised against installing the termination jumpers.
Applies to
- Models
- Precision
- Product
- Precision 500D
- Area
- X-ray R&F
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy for this GE Healthcare X-ray R&F issue are available to FSELIB subscribers — re-authored, web-first, searchable.
7-day free trial · no credit card required
Related GE Healthcare X-ray R&F issues
- Precision 500D Table Fluoro Carriage Front Bumper is not properly adjusted, allows excessive travel toward the front of the table.
- Precision 500D Aurora, dose readout does not update when an overhead Rad Tube Exposure is made, at any technic due to abnormally low mR/mAS from the Rad Overhead XRT.
- P500D Aurora with M4 Software (May 2006 Release), Dose Readout Correct on Console, Dose Printer Prints 0.0 due to a M4 Software Bug
- Precision 500D Aurora, RCIM E-stop circuit is difficult to troubleshoot because there is no overall interconnect schematic from the RCIM operator console to the Corona 2 board in the System Cabinet.
- Precision 500D Aurora system with Magic PC.
- P500D Siemens Collimator lamp voltage adjustment instructions missing due to documentation issues.