P500D Question about CAN termination jumpers for systems without an OTS or Wallstand?

Precision 500D Corona2 board does not need 120 ohm Can Bus termination

Symptom

P500D Question about CAN termination jumpers for systems without an OTS or Wallstand? There should never be any CAN termination jumpers installed for the Corona 2 with or without OTS and Wallstand. Odetta and Lightning required CAN termination jumpers, but Corona 2 based Thunder and Aurora systems do not. In fact, it was advised against installing the termination jumpers.