GE Healthcare · CTLightSpeedVCT260118327
LightSpeed 5.X (80/100kW)
Symptom
System will not scan or do tube warm-up. Gantry rotates but will not make X-ray. All indications are that the HV system is not communicating with the gantry. System will not scan or do tube warm-up. Gantry rotates but will not make X-rays. All indications are that the HV system is not communicating with the gantry.
Error message
Error: 260118327 Can not initialize Jedi generator See Long Description for complete error code listing. Fri Oct 20 19:11:03 2006 Error: 260118327 Host: orp Process: orp_subsys File: gen_exec.c Line: 1122 Fri Oct 20 19:11:02 2006 Host : On-Board Rotating Processor Ermes # : 260118327 Exception Class : Abort Severity : Secondary File : gen_exec.c Line# : 1122 Function : No System Function Reported Scan Type : Helical Scan: 8810/2/2 Can not initialize Jedi generator If this continues to appear, generator configration is mismatched, or power and commnication are not interconnected.
Applies to
- Models
- LightSpeed, VCT
- Product
- LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Area
- CT