LightSpeed 5.X (80/100kW)

Symptom

System will not scan or do tube warm-up. Gantry rotates but will not make X-ray. All indications are that the HV system is not communicating with the gantry. System will not scan or do tube warm-up. Gantry rotates but will not make X-rays. All indications are that the HV system is not communicating with the gantry.

Error message

Error: 260118327 Can not initialize Jedi generator See Long Description for complete error code listing. Fri Oct 20 19:11:03 2006 Error: 260118327 Host: orp Process: orp_subsys File: gen_exec.c Line: 1122 Fri Oct 20 19:11:02 2006 Host : On-Board Rotating Processor Ermes # : 260118327 Exception Class : Abort Severity : Secondary File : gen_exec.c Line# : 1122 Function : No System Function Reported Scan Type : Helical Scan: 8810/2/2 Can not initialize Jedi generator If this continues to appear, generator configration is mismatched, or power and commnication are not interconnected.

Applies to

Models
LightSpeed, VCT
Product
LightSpeed 5.X (80/100kW) LightSpeed 5.X (80/100kW) ~ Use with both MDAS and GDAS ~ GRE Console HP_H_G16 HP_H_M16 HP_O_M16 LightSpeed 5.X (100kW) GDAS16 with GRE Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
Area
CT

Full cause & step-by-step resolution

The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.