GE Healthcare · CTHiSpeed2300110084680001115223748718327
Scanner Hardware Stopped Scan with scan data errors
Symptom
"Scan hardware Stopped Scan" with scan data errors. "Scan hardware Stopped Scan" with scan data errors
Error message
Error: 230011008 Unable to remove scan file from scan data disks. Error: 230011008 Unable to remove scan file from scan data disks Error: 230011008 Unable to remove scan file from scan data disks. Wed May 4 16:25:48 2005 Error: 230011008 Host: gsgect2 Process: scanmgr File: sdio_xfs_io_ops.c Line: 1164 Function: Data Acquisition : SBC Processing Scan Type: Axial Scan: 2800/200/3 Scan Seq Id: 468000 Unable to remove scan file from scan data disks. Scan: CT02/2800/200/3 Scan file: /raw_data/ex2800.CT02.1115223748.718327/ov.s200a3 OS error : 140 --- Remote I/O error Scan file may be deleted manually.
Applies to
- Models
- HiSpeed
- Product
- HiSpeed QX/i - GOC2 103L.5_HSQXIM4 103L.4_HSQXIM4 102L.1_HSQXIM3 HiSpeed QX/i - GOC2 Method of visual identification Method of remote or software identification - See attachment below for a complete list of software commands and typical results. check_config List of Supported Console Types GOC2 List of softwares the product can run 103L.5_HSQXIM4 103L.4_HSQXIM4 - obsolete 102L.1_HSQXIM3 - obsolete FAQ's Table weight limit - Up to 400 lb. (180 kg) with ± .25 mm positional accuracy. 450Ib (205 kg) max weight Tube type - D3186T Current production details - See Technical Reference Manual found on the Common Document Library Tips and workarounds, where to find it - Common Document Library(CDL) . Search for the latest FMI software release. In the furnished materials will be the list of various manuals with part numbers for the various languages. Options - TBD Links to the updated Product Service Manuals, FMIs, Service Notes, Option Manuals and Operator Documents can be found online at the attached sites. Always use the most appropriate documentation for your system.
- Component
- 2339284 RIP / Motorola 2400 series board (Replaces 2197234-2) See long description for deatils 2339284 RIP / Motorola 2400 Series Card (replaces 2197234-2) Part #: 2339284 Part Type: GP CONSUMABLE - RETURNABLE Part Weight: 1.10 LBR Repairable: No Returnable: Yes Hazardous Material: N Description: PPC, 2433-1 WITH HAWK 3 CHIP FRU: Yes Status: Orderable Comments: Please consider ordering the used part when it is appropriate for your customer. It will save you substantial money. 2/3/05 (212001261) Please be sure to read the FE Assist comments for the 2370825
- Area
- CT
Full cause & step-by-step resolution
The complete cause analysis, the step-by-step remedy and the attached service documents for this GE Healthcare CT issue are available to FSELIB subscribers — re-authored, web-first, searchable.